12.21 Cycle 33 Calibration Program

The Cycle 33 calibration plans for COS continue the routine calibration and monitoring programs performed in previous Cycles. Brief descriptions of the programs are listed in Table 12.2. The full calibration plan for Cycle 33 is listed in Table 12.27. Two contingency programs introduced in Cycle 17 are included in this cycle; both FUV and NUV Detector Recovery After Anomalous Shutdown. Two contingency programs introduced in Cycle 28 are included in this cycle as well. The program COS FUV Change in Spectroscopic Sensitivity Trends is a supplement to the COS FUV Spectroscopic Sensitivity Monitor and is activated when the time-dependent sensitivity appears to be changing on short timescales. The COS FUV Characterization of Modal Gain When Changing High Voltage program is activated when the high voltage of the FUV detector is changed.  The COS FUV Split-Wavecal Monitor, which aims to improve the wavelength solution for exposures that require split wavecals, was added in Cycle 32.


Table 12.27: Cycle 33 Calibration Programs. 

Program NameProgram ID

Cycle 33 COS NUV MAMA Fold Distribution

17935

Cycle 33 COS NUV Detector Dark Monitor

17934

Cycle 33 COS NUV Spectroscopic Sensitivity Monitor

17936

Cycle 33 COS NUV Wavelength Scale Monitor

17937

Cycle 33 COS NUV Target Acquisition Monitor

18203

Cycle 33 COS FUV Detector Gain Maps

18198

Cycle 33 COS FUV Detector Dark Monitor

18197

Cycle 33 COS FUV Spectroscopic Sensitivity Monitor

18199

Cycle 33 COS FUV Wavelength Scale Monitor

18201
Cycle 33 COS FUV Target Acquisition Monitor18200
Cycle 33 COS FUV Split-Wavecal Monitor18202
Cycle 33 COS FUV Detector Recovery After Anomalous Shutdown*
17938

Cycle 33 COS NUV Detector Recovery After Anomalous Shutdown*
17939

Cycle 33 COS FUV Change in Spectroscopic Sensitivity Trends*18204
Cycle 33 COS FUV Characterization of Modal Gain When Changing High Voltage*18205

*contingency program


If your program requires calibrations beyond those described here, then you should explicitly include this calibration in your Phase I proposal.