5.10 Wavelength Settings and Ranges
Table 5.3 and Table 5.4 show the wavelength ranges recorded on the detectors for each valid combination of grating and central-wavelength setting at the nominal FP-POS=3
position (Section 5.8.2). The wavelength ranges spanned at other FP-POS
settings may be estimated using the FP-POS
step values provided in Chapter 13. Note, however, that uncertainties in the positioning of the Optics Select Mechanisms (Section 3.1.3) correspond to about half of an FP-POS
step. These wavelength ranges are subject to change as the instrumental calibration evolves. The most recent measurements are available from the COS website.
To assist users in preparing their Phase II proposals for Cycle 25 and later COS observations, a COS/FUV Wavelength Tool has been developed which allows users to visualize how the wavelength ranges covered by the allowed COS FUV settings (i.e., settings in supported mode) overlap with those of settings that have been withdrawn (i.e., not in supported mode) to preserve the lifetimes of the COS FUV detectors.
-
COS Instrument Handbook
- Acknowledgments
- Chapter 1: An Introduction to COS
-
Chapter 2: Special Considerations when Observing with COS
- • 2.1 COS FUV Detector Lifetime Positions
- • 2.2 Visit Length
- • 2.3 Central Wavelength Settings Added in Cycle 26
- • 2.4 ORIENT constraints for Extended Sources
- • 2.5 COS Observations Below 1150 Angstroms: Resolution and Wavelength Calibration Issues
- • 2.6 Time-Dependent Sensitivity Changes
- • 2.7 Spectroscopic Use of the Bright Object Aperture
- • 2.8 Non-Optimal Observing Scenarios
- • 2.9 NUV Spectroscopic Acquisitions
- • 2.10 SNAP, TOO, and Unpredictable Source Programs with COS
- • 2.11 Choosing between COS and STIS
- Chapter 3: Description and Performance of the COS Optics
- Chapter 4: Description and Performance of the COS Detectors
-
Chapter 5: Spectroscopy with COS
- • 5.1 The Capabilities of COS
- • 5.2 TIME-TAG vs. ACCUM Mode
- • 5.3 Valid Exposure Times
- • 5.4 Estimating the BUFFER-TIME in TIME-TAG Mode
- • 5.5 Spanning the Gap with Multiple CENWAVE Settings
- • 5.6 FUV Single-Segment Observations
- • 5.7 Internal Wavelength Calibration Exposures
- • 5.8 Fixed-Pattern Noise
- • 5.9 COS Spectroscopy of Extended Sources
- • 5.10 Wavelength Settings and Ranges
- • 5.11 Spectroscopy with Available but Unsupported Settings
- Chapter 6: Imaging with COS
- Chapter 7: Exposure-Time Calculator - ETC
-
Chapter 8: Target Acquisitions
- • 8.1 Introduction
- • 8.2 Target Acquisition Overview
- • 8.3 ACQ SEARCH Acquisition Mode
- • 8.4 ACQ IMAGE Acquisition Mode
- • 8.5 ACQ PEAKXD Acquisition Mode
- • 8.6 ACQ PEAKD Acquisition Mode
- • 8.7 Exposure Times
- • 8.8 Centering Accuracy and Data Quality
- • 8.9 Recommended Parameters for all COS TA Modes
- • 8.10 Special Cases
- Chapter 9: Scheduling Observations
- Chapter 10: Bright-Object Protection
- Chapter 11: Data Products and Data Reduction
-
Chapter 12: The COS Calibration Program
- • 12.1 Introduction
- • 12.2 Ground Testing and Calibration
- • 12.3 SMOV4 Testing and Calibration
- • 12.4 COS Monitoring Programs
- • 12.5 Cycle 17 Calibration Program
- • 12.6 Cycle 18 Calibration Program
- • 12.7 Cycle 19 Calibration Program
- • 12.8 Cycle 20 Calibration Program
- • 12.9 Cycle 21 Calibration Program
- • 12.10 Cycle 22 Calibration Program
- • 12.11 Cycle 23 Calibration Program
- • 12.12 Cycle 24 Calibration Program
- • 12.13 Cycle 25 Calibration Program
- • 12.14 Cycle 26 Calibration Program
- • 12.15 Cycle 27 Calibration Program
- • 12.16 Cycle 28 Calibration Program
- • 12.17 Cycle 29 Calibration Program
- Chapter 13: Spectroscopic Reference Material
- • Glossary