0.1X0.03 Aperture

Description

This slit is supported for use with the echelle gratings, and available-but-unsupported for use with the first-order gratings. During ground calibration, test observations of a mono-isotopic platinum lamp source with this slit and the high resolution echelle gratings exhibited resolving powers in excess of 200,000 – the highest available with STIS. 

Special Considerations

Observations with this slit require a prior PEAKUP. The PEAKUP is performed in two stages—first with the 0.2X0.09 aperture (or one of the small neutral density apertures), and then with the 0.1X0.03 aperture (see Section 8.3, Table 8.5).  The throughput can be particularly sensitive to degraded focus (ISR 2017-01), and specialized data reduction techniques are required to correct for fixed pattern noise in the MAMA highres format (Jenkins & Tripp 2001, ApJS, 137, 297).


Table 13.28: 0.1X0.03 Throughputs.

λ

% Throughput

1200

19.9

1800

31.4

2400

36.0

3000

36.4

3600

34.8

4200

32.8

4800

30.9

5400

28.9

6000

27.0

Projected Length (arcsec)

Aperture

MAMA (E-modes)

Width (arcsec)

0.1X0.03

0.1

0.025

Figure 13.75: 0.1X0.03 Aperture Throughput as a Function of Wavelength.