52X0.1 Aperture

Description

This slit is supported with use of all first-order gratings and is available-but-unsupported for the echelle gratings. This aperture is the 2 pixel-wide slit in the dispersion direction for the CCD first-order modes. 

Special Considerations

Observations with this slit require a PEAKUP prior to science observations. Use of this long slit with an echelle grating will cause order overlap problems (particularly for the most closely spaced orders; see Section 12.2), but for point sources the order separation may be adequate for many science programs.


Table 13.19: 52X0.1 Throughputs.

λ

% Throughput

1200

50.1

1600

60.4

2000

66.8

2400

71.6

2800

74.1

3200

75.0

3600

74.7

4000

74.1

4400

73.5

4800

72.7

5200

72.0

5600

71.3

6000

70.6

6400

70.0

6800

69.2

7200

68.3

7600

67.4

8000

66.6

8400

65.9


Projected Length (arcsec)


 

Aperture

CCD

MAMA (M-modes)

MAMA (L-modes)

Width (arcsec)

52X0.1

52.0

28.0

25.0

0.1