5.6 Generic Detector and Camera Properties

5.6.1 Full Well Depth

Conceptually, full well depths can be derived by analyzing images of a rich starfield taken at two significantly different exposure times, identifying bright but still unsaturated stars in the short exposure image, calculating which stars will saturate in the longer exposure and then simply recording the peak value reached for each star in electrons (using a gain that samples the full well depth, of course). In practice, it is also necessary to correct for a ~10% "piling up" effect of higher values being reached at significant levels of over-saturation relative to the value at which saturation and bleeding to neighboring pixels in the column begins (see WFC3 ISR 2010-10).

Since the full well depth varies over the CCDs, it is desirable to observe a rich star field with a gain that samples the full well depth (e.g. the default WFC3 UVIS gain), and for which a large number of stars saturate. Calibration and GO programs have serendipitously supplied the requisite data of rich fields observed at two different exposure times.

There is a real and significant large-scale variation of the full well depth on the UVIS CCDs. The variation over the UVIS CCDs is from about 63,000 e- to 72,000 e- with a typical value of about 68,000 e-. There is a significant offset between the two CCDs, as visible in Figure 5 of WFC3 ISR 2010-10. The calwf3 calibration pipeline uses an average value (SATURATE, in the CCDTAB file) for full well in order to flag saturated pixels in the science image data quality file.

5.6.2 Linearity at Low to Moderate Intensity

Linearity at low and moderate exposure levels is measured by comparing back-to-back exposures of NGC 1850 made in October 2009. Figure 5.14 shows the response of UVIS2, where aperture sums for stars with flux greater than about 2,000 e- in a short exposure (central pixel greater than about 350 e-) show apparently perfect linear response when compared to the counts in the same aperture in an exposure 50 times as long. However, below a level of 2,000 e- the ratio of long to short exposure counts deviates from a linear response. At a total aperture flux of about 200 e- in the short exposure, the total flux values are ~ 5% lower than expected based on scaling from the corresponding long exposure.

Figure 5.15 shows data from both UVIS CCDs for stars with short exposure aperture sums of 500 to 2000 e-. Stars near the readout amplifiers (Y pix of 0 and 4095) behave as expected i.e. the short exposure sums scaled by the difference in exposure time are equivalent to the long exposure but losses grow with distance from the amplifiers. This is consistent with losses induced by finite charge transfer efficiency in successive parallel shifts in clocking the charge packets off the CCDs.

The WFC3 team monitors the extent to which CTE losses influence faint object photometry. CTE degradation in flight has increasingly affected the measurement of unsaturated images over the years, requiring increasingly greater flux corrections with larger uncertainties. Results from the past and ongoing calibration programs are summarized in Chapter 6.

Figure 5.14: Loss of linearity at low and moderate intensity due to finite Charge Transfer Efficiency.

The upper panel shows the counts in r = 3 pixel apertures for several thousand stars in back-to-back 500s and 10s exposures plotted against each other. To better illustrate small differences, the lower panel shows the same data after binning and plotting as the ratio of long to short exposure time, where 50.0 denotes perfect linearity.


Figure 5.15: CTE Losses For Short Exposure Image for UVIS1 and UVIS2

Ratio of fluxes in short and long exposures as a function of Y position. This shows the data from the previous figure for UVIS2, plus similar data for UVIS1 plotted against y-position in the concatenated detector space. Y-values near 2050 correspond to the maximum distance from the readout amplifiers, and hence the most parallel shifts inducing CTE losses.


5.6.3 Linearity Beyond Saturation

The response of the WFC3 UVIS CCDs is linear well beyond the point of saturation. WFC3 ISR 2010-10 shows the well-behaved response of WFC3: electrons are clearly conserved after saturation -- in some locations with the need for a minor calibration, as provided in the ISR, in other regions no correction is needed. This result is similar to that of the STIS CCD (Gilliland et al. 1999), the WFPC2 camera (Gilliland, 1994), and ACS (ACS ISR 2004-01). It is possible to easily perform photometry on point sources that remain isolated simply by summing over all of the pixels into which the charge has bled.

To characterize the accuracy of point source photometry for sources in which one or more pixels have exceeded the physical full well depth, we used a dataset consisting of alternating short and long exposures of a moderate-to-rich star field, resulting in both unsaturated and saturated images for many stars. Counts were summed over an area designed to include all pixels in the core of the stellar image and all pixels that were bled into in the long exposures.

Departures from linearity were greater for UVIS1 than for UVIS2. Results for Amp A and UVIS1 are summarized in Figure 5.16 and Figure 5.17. Over a range of nearly 7 magnitudes beyond saturation, photometry remains linear to ~ 1% after a simple correction (taken from WFC3 ISR 2010-10). For Amp C on UVIS2, the response is sufficiently linear beyond saturation that no correction is required. These results were obtained when CTE was near pre-flight levels in October 2009. See Section 5.6.2 for the level of CTE loss that was already measurable at that time. CTE degradation in flight has increasingly affected the measurement of unsaturated images over the years, requiring increasingly greater flux corrections with larger uncertainties.

These results are based on the use of FLT images. The flux conservation ensured by AstroDrizzle leads to equally good results for linearity beyond saturation when comparing long and short *_drz.fits images.

Figure 5.16: Linearity Analysis for Amp A

Linearity beyond saturation for Amp A on UVIS1. The upper panel shows the ratio of counts in the long exposure divided by the counts in the short exposure multiplied by the relative exposure time; linearity would thus result in a value of unity. The x-axis shows the multiplicative degree by which a star is over-saturated in the long exposure. The middle panel shows the long exposure aperture sums versus short exposure aperture sums. The lower panel shows the peak data value in the long exposure relative to the short exposure value. The response is linear up to the value of 68,000 e- (y-axis) where the long exposure encounters saturation. Amp A data show significant deviations from a linear response for over-saturations near and beyond 10.



Figure 5.17: Long vs. Short Exposure Ratios of Linearity

Top panel shows the count rate ratio between long and short exposures for both amplifier quadrants of UVIS1 (Amps A and B) vs. the over-saturation level in the long exposure. Bottom panel shows the same data but after applying the corrections as described in WFC3 ISR 2010-10At high levels of saturation, the Long to Short count ratios are restored to unity and the scatter is much reduced.


5.6.4 Shutter Stability

The WFC3-UVIS shutter is a circular, rotating blade divided into two open and two closed quadrants (See Section 2.3.3 of the WFC3 Instrument Handbook for details). Operationally, the shutter mechanism has two distinct modes, based on commanded exposure times. At the shortest commanded exposure time of 0.5 seconds, the shutter motion is continuous during the exposure, rotating from the closed position through the open position and on to the next closed position. For commanded exposure times of 0.7 seconds and longer (0.6 seconds is not allowed), the shutter rotates into the open position, stops and waits for an appropriate amount of time, and then rotates to the closed position.

For short exposure times, detector position-dependent exposure time (shutter shading), A versus B blade shutter dependence, stability, and timing accuracy were assessed using data taken during SMOV. For a full discussion of the analysis of shutter behavior from on-orbit data see WFC3 ISR 2009-25 and WFC3 ISR 2015-12Stellar PSFs in short exposures taken with shutter B show slightly larger FWHM and lower sharpness compared to those taken with shutter A. As described in detail in WFC3 ISR 2014-09, this effect is particularly noticeable in the shortest, 0.5-sec, exposures. To facilitate the science where short exposures are necessary and a good and stable PSF is crucial, an available-mode option is available in APT to select the shutter blade A. No other systematic difference in shutter behavior (exposure time, repeatability, etc.) is found when comparing the A and B blades of the shutter. Even at the shortest exposures, the measured shutter shading does not exceed ~0.2% across the detector. The small magnitude of this effect means that no correction for shutter shading is necessary in calwf3.

The stability of shutter timing is a bit more problematic. Results are based on eleven pairs of back-to-back exposures at each commanded exposure time. For exposure times of 1.0 seconds or shorter, the rms variation in exposure time for a series of images is 1% or greater, implying possible difficulty in achieving 1% photometric accuracy. For a commanded exposure time of 0.5 seconds, the rms variation is 1.9%. For commanded exposure times of 0.7 and 0.8 seconds, the true exposure times vary by 1.5% and 1.4% respectively. At an exposure time of 1.0 second, the rms variation falls to 1.0%.

While the rms variations were all less than 2%, we observed individual exposures at each commanded exposure time that deviated by larger amounts. For the 0.5, 0.7, 0.8 and 1.0-second exposures, we found individual exposures with measured errors of 4.0%, 4.0%, 3.0%, and 2.0% respectively. This implies that exposures of 1.0 seconds or shorter may experience timing fluctuations that could compromise a goal of 1.0% or even 2.0% accuracy. This conclusion regarding shutter stability is not regarded as robust, but is most consistent with a simple and conservative interpretation of the test data.

Finally, our investigation of measured versus commanded exposure times indicated that for exposures commanded to be 0.5 seconds, the shutter was actually open for 0.48 seconds. Similarly, for exposures commanded to be 0.7 seconds, the measured exposure time was in fact 0.695 seconds. For these exposure times, the EXPTIME header keyword value is updated in the science image headers to reflect the actual (as opposed to commanded) exposure times.

5.6.5 Fringing

At wavelengths longer than about 650 nm, silicon becomes transparent enough that multiple internal reflections in the UVIS detector can create patterns of constructive and destructive interference, or fringing. Fringing produces wood-grain patterns in response to narrow-band illumination at long wavelengths, see Figure 5.18.

Figure 5.18: Fringing

Quadrant B of two ground flat fields: one affected by fringing (right) and one not affected (left). Black region of the FQ906N flat is masked to avoid areas affected by the quad filter edges; the same region of F673N is masked for consistency.


The amplitude and phase of the fringes is a strong function of the silicon detector layer thickness and the spectral energy distribution of the illumination. Fringe amplitude i.e. the contrast between constructive and destructive interference is greatest a) at the longest wavelength (high transparency allows more internal reflections) and b) for the narrowest spectral energy distributions (SEDs). For broad SEDs, interference is averaged over phase, so that the amplitude of the fringing is reduced. Thus fringing is significant for UVIS imaging data only if red narrow-band filters are used or sources with red line emission are observed.

Flat fields from ground tests (see WFC3 ISR 2008-46) have been used to estimate the magnitude of fringing effects, for a continuum light source, in the narrow-band red filters (see Table 5.3 and WFC3 ISR-2010-04). Each column lists a different metric of fringe amplitude for filters in which fringing effects could be detected in the flat-field data along with a control filter (F606W) without fringing. These metrics can best be understood by examining the histograms ( Figure 5.19) of the flat fields shown in Figure 5.18.

Table 5.3: Metrics of fringe amplitude based on ground flat fields

Filter

Distance Between Histogram Peaks

percent

Manual Peak-to Trough

percent

Quadrant

RMS Deviation

percent

Full Width At 20% Maximum

percent

F606W



A0.92.9

F606W



B

1.0

3.0

F606W



C

1.2

3.3

F606W



D

1.2

3.3

F656N


2.2 +/-1.2

A

1.4

4.9

F656N


1.7 +/- 1.1

B

1.3

4.5

F656N


1.7 +/- 1.2

C

1.6

5.1

F656N


3.2 +/- 1.3

D

1.5

5.1

F658N



A

1.3

4.4

F658N



B

1.2

3.9

F658N



C

1.4

4.4

F658N


0.9 +/- 1.1

D

1.3

3.7

F673N



A

0.9

3.1

F673N



B

1.0

3.2

F673N



C

1.3

3.6

F673N


0.5 +/- 1.1

D

1.4

4.2

F953N

17.8

19.8 +/- 1.7

A

7.6

23.7

F953N

17.3

20.8 +/- 1.8

B

7.8

24.0

F953N

14.9

17.8 +/- 1.7

C

7.1

23.5

F953N

15.0

11.5 +/- 1.6

D

6.7

22.9

FQ672N


4.6 +/- 1.3

D

2.0

7.0

FQ674N


2.4 +/- 1.3

B

2.1

7.4

FQ727N


2.3 +/- 1.1

D

1.6

5.3

FQ750N


1.2 +/- 1.1

B

1.3

4.5

FQ889N

7.8

10.0 +/- 1.3

A

3.8

13.5

FQ906N

12.5

12.2 +/- 1.4

B

5.4

18.2

FQ924N

6.9

10.1 +/- 1.4

C

4.4

16.5

FQ937N

9.4

14.2 +/- 1.4

D

5.0

18.4

Values are given in units of percentage of the normalized flat-field signal level. Each metric is described in the text and graphically represented in Figure 5.18

The fifth data column in the table is simply the root mean square deviation from the mean of the sample and is indicated by triangles with horizontal error bars in the histograms. Filters/quadrants with rms deviations greater than corresponding values for the control filter (F606W) may be influenced by fringing. The last column is full width at 20% maximum, rather than full width at 50% maximum because this metric is more effective for bimodal pixel brightness distributions in filters with strong fringing, such as FQ906N (pictured). The second data column gives the separation between histogram peaks, which can be detected in flat-field data for only the five reddest of the twelve filters affected by fringing. Squares in Figure 5.19 mark the histogram peaks. Adjacent fringes were also manually sampled, and the results reported in the second data column.

Figure 5.19: Histograms of the two flat-field samples shown in Figure 5.17Symbols correspond to fringe amplitude metrics listed in Table 5.3: rms deviation (triangles with error bars), full width at 20% maximum (circles with error bars), and bimodal histogram peaks (squares).


 

For estimating photometric uncertainties in data taken with these filters, it may be useful to consider how much the rms deviation in the flat fields departs from that of the control filter F606W. For exposure time calculations of targets affected by fringing, the full width at 20% maximum or the distance between histogram peaks may be the most useful metric. To understand how small dithers might affect photometry of targets that happen to fall on adjacent positive and negative fringes, the manually determined amplitude might be most appropriate. For sources with SEDs very similar to the calibration lamp, application of the pipeline flat fields should considerably reduce the effect of fringing on the data.